EasyManua.ls Logo

Omicron CPC 100 - Test Settings

Omicron CPC 100
382 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Voltage Transformer
5 - 9
Test Settings
Figure 5:
VTBurden test card
with test results
Navigate to the parameter fields, and enter the values according to your test
requirements:
Pressing the S
ETTINGS menu key opens the Settings page. The Settings
page allows setting the test cards individually. As a rule, do not use the Settings
page but the Device Setup tab in the Options view (see ”Device Setup” on
page 2-33) to set the test cards. For more information, see ”Settings Page” on
page 2-28.
Vsec: nominal secondary voltage with the option "value x 1/
3"
f: output frequency
Auto: select check box for automatic test (default), clear for manual
test (refer to page 5-10).
Vtest: secondary injection voltage from 130V AC output
Note 1: If a current clamp is used to measure Isec, select the
corresponding check box on the test card, and specify the current
clamp device at the Options tab Device Setup.
Note 2: Due to cross-talk between the measuring inputs V1AC and
V2AC, we suggest not to connect a current clamp to the input
V2AC.
Therefore, use a current clamp with current output.

Table of Contents

Other manuals for Omicron CPC 100

Related product manuals