Getting Started with Quick
3 - 11
Trigger Settings
A trigger is the occurrence of a selected event, for example, a binary trigger is
the first change of the state at the binary input.
Trigger on:
Selection of trigger event
no trigger (default)
binary trigger = CPC 100 input BIN IN is monitored for a trigger event
trigger = first measurement value (m1) > set threshold value
trigger = first measurement value (m1) < set threshold value
trigger = second measurement value (m2) > set threshold value
trigger = second measurement value (m2) < set threshold value
overload = the trigger event is an overload condition that occurs at
the selected output.
What is an "Overload" trigger?
At the 800A AC output, an overload trigger is a condition in which the
set current cannot be reached any longer, for example, because of
an opening contact or circuit breaker.
Note: Current values
< 50A do not initiate an "Overload" when the
current circuit opens. Therefore, if you use the trigger condition
"Overload", chose a nominal current value of
50A.
Quick differentiates two "Overload" trigger conditions:
1. the occurrence of an overload (as described above)
2. the clearing of an overload condition (clearing is delayed by
100 ms to debounce).
1 2
3
4
5
1

Table of Contents

Other manuals for Omicron CPC 100

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Omicron CPC 100 and is the answer not in the manual?

Omicron CPC 100 Specifications

General IconGeneral
BrandOmicron
ModelCPC 100
CategoryTest Equipment
LanguageEnglish

Related product manuals