Transformer
6 - 9
After then, pressing the A
DD T AP menu key repeatedly adds more taps with a
step calculated from the values for the first two taps. After adding all taps, press
the M
AIN PAGE menu key to transfer the data to the main page.
Note: After the transformer’s ratio was specified on the Settings page for one
phase, by pressing the S
AVE AS DEFAULT menu key you can save the data.
Then a new test card for the next phase will be inserted with the transformer’s
ratio preset.