Introduction
2 - 7
They provide an output voltage of either
• 0 ... 0.5kV with a maximum current of 5A, or
• 0 ... 1kV with a maximum current of 2.5A, or
• 0 ... 2kV with a maximum current of 1.25A.
For details refer to ”High-Current and High-Voltage Outputs” on page 16-5.
Both the voltage of this output and the current that flows from this output are
measured internally by CPC 100, and displayed on the respective test card.
All outputs are overload and short-circuit-proof and protected against external
high-voltage signals as well as over-temperature.
Note: The length of time a high current or a voltage can be applied may be
limited due to a high temperature occurrence of the internal
transformer windings and/or heat dissipaters. This high temperature
occurrence may even be increased if CPC 100 is operated in a hot
environment or is exposed to direct sunlight.
It is highly recommended not to exceed the typical duration time t
max
specified for each current and voltage output (refer to ”CPC 100
Outputs” on page 16-5). If you do, and CPC 100 develops too high an
internal temperature, the outputs are shut off automatically.