CPC 100 V1.41
2 - 18
• Test procedure
A test procedure contains multiple test cards.
The composition of such a test procedure and the settings of all single test
cards can be freely defined by the user. Within a test procedure, each test
card and its associated test is executed individually in a user-defined order.
Note: With the CPC 100 V0, the number of test cards in one test
procedure should be limited to 15 to avoid memory problems. The CPC 100
V1 allows using more test cards in one test procedure but we recommend not
to use more than 15 test cards to keep the tests clearly structured.
• Test procedure default
A test procedure with all of its test cards and specific settings - but without
test results - can be made the test procedure default.
When the CPC 100 software starts, it automatically loads the test procedure
default.
Initially, that is, as long as no user-defined test procedure default is created,
the the CPC 100 software loads one empty Quick test card.
• Report
For archiving or reporting purposes, or later processing, a test procedure with
all of its test cards, specific settings and - after the test was run - test results
and assessments can be saved. It is then considered a report.
Such a report can later be opened any time in CPC 100 File Operations
menu.
• Test procedure template
A report can be copied/pasted to be a test procedure template.
In this process, the settings are preserved, however, the test results are
deleted.