Installing Remote Devices 7 Installation Instructions
145
NOTE The Insertion Tool should be set to cutting mode &= on.
Figure 23 Connections for BIS Module
4 Use a small screwdriver to connect the two drain wires to the PCB, see the wiring schematic in
Figure 23.
5 Slide the PCB back on to the metallic mounting flange.
6 Use screws to fasten the mounting flange to the wall.
NOTE US version only: Fasten the rectangular wall-mounting plate to the wall. Attach the mounting flange to
the wall-mounting plate.
7 Mount the plastic cover. The plastic cover consists of two pieces:
– Frame
– Angled cover
Put the frame over the mounting insert and the PCB. Place the angled cover on top and fasten
with two screws.
8 Connect the monitor and the measurement server to the wall installation.
9 Perform the following tests as described in the Test and Maintenance section of this manual:
– Power-on test blocks
– Safety test blocks
– ECG Sync Performance Test