EasyManuals Logo

Philips IntelliVue MP70 Service Manual

Philips IntelliVue MP70
278 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #195 background imageLoading...
Page #195 background image
Checking and Calibrating the Anesthetic Gas Module 9 Anesthetic Gas Module
195
Before performing a Span calibration check, you must first perform:
• Performance Leakage Check.
• Performance Diagnostic Check.
• Performance Flowrate Check .
• Zero Calibration Check .
• Barometric Pressure Calibration Check .
• Ensure that there is enough gas in the calibration gas bottle.
• Check tubing assembly.
Figure 40 Span Calibration Equipment including Gas Canister and Spray Valve
CAUTION Ensure that the room you are working in is well-ventilated, and that the Philips M1026A Anesthetic
Gas Module exhaust is properly connected to the gas scavenging system.
1 In the Setup Gas Analyzer menu select Service Window.
2 Select Calibration to access the Gas Analyzer Calibration window.
3 Select the Select Cal Agent item from the Setup Gas Analyzer menu.
4 Pre-select the agent that is being used during calibration. If Halocarbon 22 is in use, select
Subst..
5 Connect the calibration gas bottle, the reservoir bag and the sample line as shown in Figure 40,
"Span Calibration Equipment including Gas Canister and Spray Valve".
6 Wait until the GA OCCLUSION INOP appears on the monitor. Now wait for another 10 seconds
to let the Anesthetic Gas Module completely evacuate the reservoir bag.
7 Now fill the reservoir bag with gas.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Philips IntelliVue MP70 and is the answer not in the manual?

Philips IntelliVue MP70 Specifications

General IconGeneral
BrandPhilips
ModelIntelliVue MP70
CategoryMedical Equipment
LanguageEnglish

Related product manuals