9 Anesthetic Gas Module Repairing the Anesthetic Gas Module
240
Now perform the performance checks described in the Test and Inspection Matrix on page 262.
Figure 62 Removing the Sample Cell Pneumatic Connections
Figure 63 Removing the Sample Cell Cover