Repairing the Anesthetic Gas Module 9 Anesthetic Gas Module
263
Performance
Flowrate Check
Perform Flowrate Check. Document the
actual flowrates.
Flowrates M1026A #A02/#A05:
Purge =
x1 (310 +/- 15 ml/min)
Measurement Path =
x2 (labelled value +/- 3 ml/min)
Normal =
x3 (132 - 170 ml/min)
PF: P/x1/x2/x3 or
PF: F/x1/x2/x3
where P=Pass and
F=Fail
Performance
Diagnostic Check
Perform the Diagnostic/Error Check.
Does the status of each subassembly display
as "OK" in
Gas Analyzer Diagnostic
window?
The subassemblies are:
- Meas. Assembly
- Meas. Opt. Path
- Agent-ID
- O2 Assembly
- Main PCB
- Power Supply
- Pneumatic System
- Operat. Temperature
Expected answer is "yes".
If so, Error/Diagnostic check is
passed.
PD:P or
PD:F
where P=Pass and
F=Fail
Performance Zero
Calibration Check
Does the status of each channel display as
"OK" in the
Gas Analyzer
Calibration
window after zero
calibration?
The channels are:
- Press
- O2
- CO2
- N2O
- Subst.
Expected answer is "yes".
If so, zero calibration check is
passed.
PZC:P or
PZC:F
where P=Pass and
F=Fail
Test Bloc k Na m e Test or Inspection to be performed Expected Test Result What to Record
on Service
Record