Repairing the Anesthetic Gas Module 9 Anesthetic Gas Module
265
Safety Step 1
Protective Earth.
See Safety Test Appendix for details / S (2).
With mains cable:
Maximum impedance = x1 (<=
100 mOhms)
S:P/x1/x2/x3/x4
or
S:F/x1/x2/x3/x4
where P=Pass and
F=Fail
Step 2
Enclosure Leakage Current - Normal
Condition.
See Safety Test Appendix for details / S (4).
Maximum leakage current = x2
(<= 100 uA)
Step 3
Enclosure Leak age Current - S.F.C. Open
Supply.
See Safety Test Appendix for details / S (5).
Maximum leakage current = x3
(<=500uA)
(<= 300 uA, for US and/or UL
devices)
Step 4
Enclosure Leakage Current - S.F.C. Open
Earth.
See Safety Test Appendix for details / S (6).
Maximum leakage current = x4
(<=500uA)
(<= 300 uA, for US and/or UL
devices)
Test Bloc k Na m e Test or Inspection to be performed Expected Test Result What to Record
on Service
Record