LIFEPAK 20/20e Defibrillator/Monitor Troubleshooting
Using the Service/Status Features (continued)
Device Log Select DEVICE LOG on the SERVICE/STATUS submenu to view essential device
characteristics, such as when the operating software was installed, and
accumulative device operations, such as the shock count.
The device log data includes the following information:
Data Description
Manufacturing date The date when the device was manufactured,
specifically, when the operating software was loaded.
Fault Messages Indicates whether there are any error codes stored in
the Service Log (refer to Processing Service Codes).
Power Cycle Count The number of times the device has been powered on.
Pacing Count Total pacing pulses delivered by the device.
Shock Count Total times the device defibrillation capacitor has been
charged.
Power On Time Total device power-on time.
Printer On Time Total printer running time.
SpO2 Operating Time Total SpO2 running time.