LIFEPAK 20/20e Defibrillator/Monitor Replacement Procedures
Assembly Tools List
Tools List The suggested list of tools for the LIFEPAK 20/20e defibrillator/monitor
replacement procedures is as follows:
■ Static-dissipative mat and wrist strap
■ Anti-static rack and/or conductive bags
■ Capacitor discharge tool (information about Capacitor Discharge Tool)
■ Torque screwdriver(s) - required torque settings are 6.8 and 10.0 in-lb.
■ Point 1 power drive bit (P1)- Phillips tip for 4-40 screws (shaft length of 2”)
■ Point 2 power drive bit (P2)- Phillips tip for 6-32 screws (shaft length of 2”)
■ 3/32” Allen Driver
■ 1/4” Nut Driver
Torque Measurements Values
Torque in-lb 6.8 10.0
Torque in-oz 108.8 160
Torque cNm 77 113
9-12