LIFEPAK 20/20e Defibrillator/Monitor Replacement Procedures
Boardstack (continued)
Boardstack
Reassembly
Note: If there are indications that the plastic isolation mounts are cracking, then
use the boardstack repair kit MIN 3202718-021 as part of the
reassembly process.
To reassemble the boardstack:
1. Install the A04 Therapy PCB.
2. Install the A01 System PCB.
3. Install the A06 OEM/A02 PP PCB.
4. Install the A10 SpO2 Module, if removed.
5. Install the boardstack assembly.
6. Install the front case.
7. Install the top case.
8. Install the A07 Battery.
9. Review the labels parts list and install new labels.
10. Complete the LIFEPAK 20 Device PIP.
Parts Lists Interconnect System View Therapy View Main Assemblies