LIFEPAK 20/20e Defibrillator/Monitor Modes of Operation
Service Mode (continued)
Resetting the
Maintenance Prompt
After completing scheduled maintenance, reset the maintenance prompt counter
to clear the
MAINTENANCE DUE message and begin the count for the next
scheduled maintenance.
To turn off or reset the scheduled maintenance prompt:
1. Display the
SERVICE menu.
2. Select
MAINT PROMPT. The SERVICE/MAINT PROMPT menu appears, showing
the current prompt date for scheduled maintenance.
3. Select
RESET. The prompt date is revised to the next scheduled maintenance
date.
4. Turn the device OFF.