LIFEPAK 20/20e Defibrillator/Monitor Troubleshooting
Device Printed Circuit Board Replacement
Anytime a printed circuit board (PCB) is replaced, then reloading of device
system software to the current version is required.
The device system software are multiple software programs maintained on the
circuit boards within the LIFEPAK 20 defibrillator/monitor, know as a device
software set. This device software set is documented as a single software
version and part number. Updating the software will maintain the device at a
known released device system software set.
Device Software set can by install by using the LIFENET System Asset
management tools or by contacting your local Physio-Control service or sales
representative.