LIFEPAK 20/20e Defibrillator/Monitor Replacement Procedures
Boardstack (continued)
Boardstack
Disassembly
To disassemble the boardstack:
1. Remove the A07 Battery.
2. Remove the top case.
3. Remove the front case.
4. Remove the boardstack assembly.
5. Remove the A10 SpO2 Module (only if it is being replaced).
6. Remove the A06 OEM/A02 PP PCB.
7. Remove the A01 System PCB.
8. Remove the A04 Therapy PCB.