LIFEPAK 20/20e Defibrillator/Monitor Modes of Operation
Setup Mode (continued)
Creating a Passcode
ARCHIVES ACCESS – Set the device to any of the following protocols (refer to
the table above):
1. Allow unlimited access to archives mode and allow records to be deleted.
2. Require a password to enter archives mode, but allow records to be
deleted.
3. Allow unlimited access to archives mode, but require a password to delete
records.
4. Require a password to enter archives mode and delete records setup
MODE – Create a new passcode to access the SETUP menu. The ENTER
PASSCODE overlay appears with the first digit highlighted. Rotate the
SPEED DIAL to select digits.
ARCHIVES MODE – Create a passcode to enter archives mode. The ENTER
PASSCODE overlay appears with the first digit highlighted. Rotate the SPEED
DIAL to select digits.
DELETE RECORDS – Create a passcode to delete records in archives mode.
The
ENTER PASSCODE overlay appears with the first digit highlighted. Rotate
the
SPEED DIAL to select digits.
4-10