LIFEPAK 20/20e Defibrillator/Monitor Preventive Maintenance
Storing the Device
When not in use, or during long periods of storage, connect the device to ac
power. If this is not possible, fully charge the batteries at an ambient room
temperature, not to exceed 25° C (77° F), prior to storage and before use.
Note: Do not store or ship the device without turning off the AC Loss Alert
alarm.
AC Loss Alert Alarm
The device is equipped with an alarm that beeps when the device is turned off
and not connected to an ac power source. The alarm can be configured to beep
at 5-, 15-, or 30-minute intervals, or it can be turned off. The default setting is
15 minutes.
To set or disable the alarm:
1. Display the
SETUP menu.
2. Select
GENERAL from the SETUP menu.
3. Select
AC LOSS ALERT from the SETUP/GENERAL submenu.
4. Select
5 MINUTES, 15 MINUTES, 30 MINUTES, or NEVER ALERT to set or turn off
the alarm.