LIFEPAK 20/20e Defibrillator/Monitor Replacement Procedures
Saving the Setup Configuration (continued)
2. Display the SETUP menu on both devices.
3. Click
SEND CONFIG in the SETUP menu on the spare device. The SEND
CONFIG overlay appears.
4. Select
SEND and press the SPEED DIAL. The setup configuration transfers
back to the repaired device.
5. Turn both devices OFF.
Printing the Setup
Configuration
To print the setup configuration:
1. Display the
SETUP menu.
2. Select
PRINT DEFAULTS. The printer prints the device setup configuration.
Save this printout for future reference.
3. Turn the device OFF.
4. Make the necessary repairs.
5. Turn the device ON and display the
SETUP menu.
6. Using the printout, check the settings in each menu and revise as necessary
to match the printout.
7. Turn the device OFF.