LIFEPAK 20/20e Defibrillator/Monitor Replacement Procedures
Saving the Setup Configuration (continued)
Transferring the Setup
Configuration
To transfer the setup configuration to a spare device:
1. With the power OFF on both devices, connect the two devices using a
configuration transfer cable (P/N 3202447) between the device system
connectors.
2. Display the
SETUP menu on both devices.
3. Select
SEND CONFIG in the SETUP menu on the device to be repaired. The
SEND CONFIG overlay appears.
4. Select
SEND and press the SPEED DIAL. The setup configuration transfers to
the spare device.
5. Select
PRINT DEFAULTS in the SETUP menu on the device to be repaired. The
printer prints the device setup configuration. Save this backup printout for
possible future reference.
6. Turn both devices OFF.
Restoring the Setup
Configuration
To restore the setup configuration by transferring it back to the repaired device:
1. Connect the spare device (with the saved setup configuration) to the
repaired device using a configuration transfer cable (P/N 3202447) between
the device system connectors.
(Continued on next page)