LIFEPAK 20/20e Defibrillator/Monitor Device Description
Functional Description (continued)
A15 Elastomer Keypad
The A15 Elastomer Keypad displays the common device controls (those not
available using the
SPEED DIAL). The number of keys on this keypad varies,
depending on the features installed in a specific device.
A19 AC Input Power
Filter
The A19 AC Input Power Filter provides input current overload and
electromagnetic interference (EMI) protection for the device. The filter is a potted
module containing passive filter elements (inductors and capacitors), with in-line
fuses in both the line and neutral leads. The A19 AC Input Power Filter is
designed to meet the safety requirements in IEC 60601-1.
W01 Therapy
Connector Assembly
The W01 Therapy Connector Assembly provides a patient connection port
used for delivery of either defibrillation or pacing therapeutic energies. The
standard and premium models allow the attachment of all available electrode
accessories, including QUIK-COMBO pacing/defibrillation/ECG electrodes,
external standard paddles (with built-in pediatric paddles), and internal paddles
with discharge control. The W01 Therapy Connector mounts directly to the
bottom case and the wire harness plugs directly into the A04 Therapy PCB at
J13 and J14. The therapy connector protrudes through a hole in the front case to
provide user access for connecting the various external cable options.
Note: The device supports all existing LIFEPAK 12 defibrillator/monitor
accessories (including external adult paddles with posterior
attachments).