LIFEPAK 20/20e Defibrillator/Monitor Preventive Maintenance
Maintenance and Testing Guidelines
Periodic maintenance, inspection, and testing of the device will help prevent
possible electrical and mechanical problems. Refer to the Operator Checklist in
the operating instructions for additional items.
The following table shows the schedule for preventive maintenance activities.
For items that should be replaced at regular intervals, refer to scheduled
replacement Items shown below.
Activity As Needed Scheduled
Performance inspection procedure (PIP) X Annually
Test and calibration procedures (TCP) X
Exterior physical inspection X
Interior Physical Inspection X
Exterior cleaning X
Interior cleaning X
Coin battery replacement 5 years
A07 battery replacement (service life
minimum of 2 years)
X
7-2