LIFEPAK 20/20e Defibrillator/Monitor Preventive Maintenance
Cleaning (continued)
Internal Cleaning
Procedure
WARNING!
Shock hazard. The energy storage capacitor carries high voltage. Remove
the battery and discharge the capacitor before handling.
CAUTION!
Possible case damage. Do not clean any part of this device or accessories
with bleach, bleach dilution, or phenolic compounds. Do not use abrasive
or flammable cleaning agents. Do not attempt to sterilize this device or any
accessories unless otherwise specified in accessory operating
instructions.
Clean the interior of the device as described below.
1. Brush interior surfaces and parts with a nonmetallic, soft-bristle brush.
2. Remove loosened dirt and dust using a dry, low-pressure compressed air
(60 psi) or vacuum cleaner.
3. Wipe metal surfaces with a soft, nonabrasive cloth that has been dampened
with isopropyl alcohol.
7-6