LIFEPAK 20/20e Defibrillator/Monitor Troubleshooting
Troubleshooting Chart
Area Observed Symptom Suggested Corrective Action
Physical Inspection Loose or broken hardware Locate and tighten or replace loose items.
Locate and replace broken components.
Evidence of dirt, fluids, or
foreign objects
Perform external cleaning.
Damaged keypad or labels Replace elastomer keypad.
Replace product identification label.
Replace explosion/hazard label.
Replace operating instruction label.
Power Off Device beeps when turned
off
Connect device to ac power source.
Disable AC Loss Alert alarm.
6-4