LIFEPAK 20/20e Defibrillator/Monitor Replacement Procedures
Front Case (continued)
W02 Speaker Assembly
Removal
Note: The following assemblies must be removed before beginning this
disassembly:
– Top case
– Front case
To remove the W02 Speaker Assembly:
1. Disconnect the speaker cable from the W25 Speaker Harness Extension
Cable connector. Refer to the W25 Speaker Harness Extension Cable
removal and installation instructions for more information.
2. Remove and discard the two 4-40 0.312 screws (173) from the speaker
mounting bracket (41), and remove the speaker mounting bracket from the
front case.
3. Remove the W02 Speaker Assembly from the front case.
Parts List Interconnect Front Case View Rear View Main Assemblies