Page 68
STAT
IM
2000/2000S Service Guide
96-108639 Rev 3.0
3. Troubleshooting Cycle Faults
STATIM 2000/2000S/2000 G4 Service Guide
CF
No.
Unit
type
Software revision
Description of Fault
Probable Cause (or causes) of Fault
S Non
S
1.xx/
2.xx
4.xx 5.xx 6.xx/
7.xx
1
The cassette temperature failed to reach 95ºC
within a time-out period.
% Large cassette leak
% Overloaded cassette
% Poor pump delivery
% Blown thermal fuses
% Contaminated steam generator
% Failed steam generator triac
% Steam generator fault (open circuit).
2
The cassette temperature failed to increase from
95°C to 100°C within a time-out period.
% Major steam leak
% Overloaded cassette
% Incorrect chamber calibration
3
The cassette has failed to pressurize and
achieve a temperature of 110ºC within a time-
out period.
This may be caused by a
% Faulty cassette seal
% Damaged cassette
% Steam circuit leak (tubing)
% Incorrect probe alignment
% Sticking solenoid valve (open)
% Check valve leak
% PRV leak
Check Vref is 2.520V± 0.001V (all software revisions
OTHER THAN revision 6.xx)
4
The cassette has failed to achieve sterilization
conditions within a timeout period.
This may be caused by a
% Faulty cassette seal
% Damaged cassette
% Steam circuit leak (tubing)
% Incorrect probe alignment
% Sticking solenoid valve (open)
% Check valve leak
% PRV leak
Check Vref is 2.520V± 0.001V (all software revisions
OTHER THAN revision 6.xx)