EasyManuals Logo

Tektronix Keithley 4200A-SCS User Manual

Tektronix Keithley 4200A-SCS
324 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #120 background imageLoading...
Page #120 background image
Section 6: Using a Model 82 C-V System Model 4200A-SCS Prober and External Instrument Control
6-12 4200A-913-01 Rev. A December 2020
4. Enable or disable cable compensation: Use the frequency and range fields to either disable (0) or
enable (1) cable compensation for the test frequencies and ranges. The following figure shows
cable compensation enabled for all ranges and test frequencies.
Figure 88: CableCompensate82 user module
5. Select Run to execute the test.
6. Follow the instructions on the dialog boxes, which will guide you through the cable compensation
process. The three basic dialog boxes are shown below.
â–ª Measure offset: An open circuit measurement is required. Open the circuit as close to the
DUT as possible.
â–ª Measure capacitance source: Connect a capacitance source in place of the DUT. Note that
the value in the dialog box corresponds to a calibration value you entered in Enter and save
capacitance source values (on page 6-10). Connect the capacitance source as close to the
DUT as possible.
â–ª Compare readings: Compares the measured value to the calibration (nominal) value you
entered. The two readings should be fairly close. If they are not, the wrong capacitance
source may have been connected or an open circuit condition occurred. In that case, click
Cancel to abort the cable compensation process.

Table of Contents

Other manuals for Tektronix Keithley 4200A-SCS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley 4200A-SCS and is the answer not in the manual?

Tektronix Keithley 4200A-SCS Specifications

General IconGeneral
BrandTektronix
ModelKeithley 4200A-SCS
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals