Operating the Autosampler 100
4
5. At the Run Standards? field, select Yes (default).
6. Move the cursor to the Vials Between field. Enter the number of sample vials to run
between standards (1 to 48 or 1 to 96, depending on carousel type).
7. Press the down-arrow key to move the cursor to the Vial field. Enter the following
parameters for each standard vial. You can enter up to five standard vials for each
Auto Run step.
Running Auto Standards
To start the Auto Run:
1. When programming is complete, press Step Function, then Auto Page to return to
the Edit: Auto Run screen.
2. Press Start Auto. The Auto Standards routine runs as part of each programmed
step. A message appears indicating that Auto Standards is running. You cannot
change screen settings of the currently running step.
Turning Off Auto Standards
To turn off Auto Addition on the Edit Auto Addition screen, select No at the Run Addition?
field. The information programmed for Auto Standards is saved, but Auto Standards do not
run.
When Auto Standards are turned off, the Stds notation is not displayed on the Auto Run
screen.
Note: If you turn off Auto Standards on the Edit: Step Function screen (by setting the
Option field to None), you delete the information programmed for Auto Standards.
4.6.3.5 Example Auto Standard Run
In an Auto Standards run, standards always run at the beginning of a step, and may run
during or at the end of a step.
Whether standards run during or at the end of a step is determined by the number of vials
in a step and the number of vials specified to run between standards.
Parameter Description/Range
Vial Carousel number, 1 to 96
Inj Vol 0.0 to 200.0 µL or 0.0 to 2000.0 µL
*
*. Dependent on sample loop installed in Waters 717plus Autosampler
# Inj Number of injections for standard vial, 1 to 99
Run Time Total run time from injection to Inject Stop
signal, 0.0 to 999.9 min