Specifications 161
A
Appendix A
Specifications
This appendix includes information on:
• Operational specifications
• Mechanical specifications
• Environmental specifications
• Power requirements
• Communications
Table A-1 Operational Specifications
Condition Specification
Injection Volume Range 0.1 µL to 2000 µL in 0.1-µL Increments
Linearity (1 to 100 µL) > 0.99 Coefficient of deviation
Carryover < 0.1 %
Bandspreading < 10 % Contribution for all solvent systems
Precision (typical, 5 to 50 µL) < 0.5 % RSD
System Operating Pressure 4500 psig (31027 Kpa) maximum continuous
operating pressure 5000 psig (34475 Kpa)
maximum surge for <30 seconds
Wash Flow, Solvent Rate > 0.4 mL per injection
Minimum Sample Required 7 µL using low-volume inserts
Sample Compartment-Temperature
Control
(Optional Kit Required)
4 to 40 °C, + 3 °C, set in 1-°C Increments; max.
time to reach set point 1.5 Hr. (*low setpoint is
derated above 27 °C ambient temperature)
Advanced Sampling Routines Priority Samples, Auto Standards, Auto Addi-
tion, and Auto Transfer