EasyManua.ls Logo

Waters 717plus - Auto Transfer Screen

Waters 717plus
181 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operating the Autosampler 108
4
Note: For instructions on setting Needle Position and Purge, see Section 4.6.2,
Programming Step Functions.
2. From the Edit: Step Function screen (Figure 4-11
), press the down-arrow key to
move the cursor to the Option? field.
3. Press the right-arrow key to move the cursor to the Trns field (Figure 4-11
). The
Trns Page key appears on the screen.
4. Press Trns Page to display the Edit: Auto Transfer screen (Figure 4-17
).
Figure 4-17 Auto Transfer Screen
5. At the Run Transfer field, select Yes (default).
6. Move the cursor to Transfer Sequence? field. Select a transfer sequence:
Step Transfers one time at the beginning of a step. The solution is transferred
from the specified vial into only the first sample vial in the step.
Vial (default) Transfers one time for each vial in the step. The solution is
transferred to each sample vial before the first injection of each vial.
Inj Transfers before every injection from every vial. Use this method for
multiple (progressive) dilutions of samples.
Note: Section 4.6.5.3,
Example Auto Transfer Run, includes examples of each
transfer sequence.
IDLE
Edit: Auto Transfer 1 of 1 Trns
Editing Step:
Run Transfer? Yes No
Transfer Sequence? Step Vial Inj
Dispense Needle (%) 25 50 75 100
Vial
1
2
Trans
Vol
10.0
20.0
Step
Function
Delete
Table
Delete
Line
#
Mix
1
2
Mix
Vol
10.0
20.0
Delay
Time
1.0
1.0
Purge
Time
1.0
1.0

Table of Contents

Related product manuals