Operating the Autosampler 110
4
Running Auto Transfer
To run the Auto Transfer routine:
1. When programming is complete, press Step Function, then Auto Page to return to
the Edit: Auto Run screen.
2. Press Start Auto to start the run. The Auto Transfer routine runs as part of the Auto
Run. A message appears indicating that Auto Transfer is running. You cannot
change screen settings of the currently running step.
Turning Off Auto Transfer
To turn off Auto Transfer on the Edit Auto Transfer screen, select No at the Run Transfer
field. The information programmed for Auto Transfer is saved, but Auto Transfer does not
run.
When Auto Transfer is turned off, the Trns notation is not displayed on the Auto Run
screen.
Note: If you turn off Auto Transfer on the Edit: Step Function screen (by setting the Option
field to None), you delete the information programmed for Auto Transfer.
4.6.5.3 Example Auto Transfer Run
Transfer Before Step
If you select Step as the Transfer sequence, the 717plus transfers a solution to the sample
vial one time at the beginning of a step. The solution is transferred from the specified
transfer vial into the first sample vial in the step.
Figure 4-18
illustrates a two-step Auto Run in which the transfer takes place from vial 1
(common vial) to the first vial of each step (vials 2 and 10).