Calibrating the Seal Pak 61
3
Figure 3-5 Path to Auto Zero Pressure Page
TP01943
Menu Page
Menu: Diagnostic
Menu: User
Maintenance
Menu: Calibration
Diag
Page
UserMaint
Page
Cal
Page
AutoZero
Page
DIAG
Diagnostic: Auto Zero Pressure
Ambient pressure with no correction
must be between 90 psi anf 750 psi.
Otherwise, manuallu adjust transducer.
Autozero
Pressure
Cal
Page
Pressure Correction
Pressure (psi)
72
0
AutoZero
Pressure