Configuring the Autosampler 67
3
Configuration parameters are shown in Table 3 -2 .
Table 3-2 Configuration Parameters
Parameter
Configuration
Page
Description
Purge In Idle? 1 Enables purging every 60 minutes while the 717plus
Autosampler is idle. See Section 4.8.5,
Using Buff-
ered Mobile Phase.
Vials Between Purges 1 Enables automatic purging during an auto run. See
Section 4.8.4,
Purging During an Auto Run.
Halt Bad Comp
Check?
1 Prevents the next injection when a compression check
(auto purge) fails. See Section 4.8.4,
Purging During
an Auto Run.
Report Delay 1 Delays the next injection to allow a data integrator to
complete the processing of the previous run. See
Section 3.6.2,
Setting Parameters for LC System
Operation.
Hold Input? 2 Enables other LC components to stop the injector.
See Section 3.6.2,
Setting Parameters for LC System
Operation.
Hold Contact Sense? 2 Sets polarity of signal arriving at the terminal strip.
See Section 3.6.2,
Setting Parameters for LC System
Operation.
Inject Start Pulse 2 Sets length of pulse used to signal other LC compo-
nents that an injection occurred. See Section 3.6.2,
Setting Parameters for LC System Operation.
Inject Stop Pulse 2 Sets length of pulse used to signal other LC compo-
nents that the run has ended. See Section 3.6.2,
Setting Parameters for LC System Operation.
Temp Setpoint 2 Sets temperature for sample compartment
heater/cooler. See Section 4.8.1,
Temperature
Control.
Temp Control? 2 Enables temperature control for units with
heater/cooler. See Section 4.8.1,
Temperature
Control.
Draw Rate 3 Sets syringe draw rate. See Section 4.8.3, Injecting
Viscous Samples.
Syringe Volume? 3 Set equal to installed syringe size. See Section 4.8.2,
Microbore/Semiprep Injections.