EasyManua.ls Logo

Waters 717plus - 3.6.2 Setting Parameters for LC System Operation

Waters 717plus
181 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Startup and Configuration 68
3
3.6.2 Setting Parameters for LC System Operation
To communicate with other LC system components:
Set Report Delay for data integrators
Set terminal strip signal parameters for Hold, Injection Start, and Injection Stop
Set IEEE-488 address
Setting Report Delay
The Report Delay provides time for the data integrator to process data from a run before a
new injection begins.
If a data integrator is attached to the LC system, set the delay as follows:
1. From the 717plus Main Menu, press Config Page.
2. On Configuration Page 1 (Figure 3-9
), highlight the Report Delay field and enter the
delay time in minutes.
Loop Volume? 3 Set equal to installed sample loop size. See
Section 4.8.2,
Microbore/Semiprep Injections.
Installed IEEE
Address
4 Displays current address. See Section 3.6.2, Setting
Parameters for LC System Operation.
Desired IEEE Address 4 Sets new IEEE address. See Section 3.6.2, Setting
Parameters for LC System Operation.
Table 3-2 Configuration Parameters (Continued)
Parameter
Configuration
Page
Description

Table of Contents

Related product manuals