Troubleshooting Procedures 159
6
Figure 6-3 Accessing the Auto Zero Pressure Page
TP01943
Menu Page
Menu: Diagnostic
Menu: User
Maintenance
Menu: Calibration
Diag
Page
UserMaint
Page
Cal
Page
AutoZero
Page
DIAG
Diagnostic: Auto Zero Pressure
Ambient pressure with no correction
must be between –90 psi anf 750 psi.
Otherwise, manuallu adjust transducer.
Autozero
Pressure
Cal
Page
Pressure Correction
Pressure (psi)
–72
0
AutoZero
Pressure