Making a Stat Run 117
4
Figure 4-23 Edit Stat Run Screen
Entering Stat Run Parameters
Note: You must enter a value for each parameter. For parameter ranges, see
Section 4.6.1,
Programming Auto Run.
To program a Stat Run:
1. Enter the first vial of stat sample sequence in the From Vial field.
2. Enter the end of sequence of stat vials in the To Vial field. If you have only one
sample, enter the vial number again in the To Vial field.
3. Enter the number of injections made from each vial in the # Inj field.
4. Enter the volume in microliters for each injection in the Inj Vol field.
5. Enter the length of time from start of injection until the End-of-Run signal and the
start of the next injection in the Run Time field.
Starting a Stat Run
After programming the stat parameters, press Start Stat to begin the run.
Note: The stat run does not begin until the run time for the last auto run injection ends.
IDLE 22
i
Edit: Stat Run
Step
1
Main
Page
Start
Stat
From
Vial
45
To
Vial
48
#
Inj
1
Inj
Vol
5.0
Run
Time
10.0
End of Table