Operating the Autosampler 104
4
Figure 4-15 Auto Addition Screen
5. From the Run Addition? field, select Yes (default).
6. Move the cursor to the Delay Time field. Enter the time to delay before injection (0.0
to 99.9 minutes). Default is 1.0 minutes.
7. Move the cursor to the Common Vial field. Enter up to five common vials.
Running Auto Addition
To start the Auto Run:
1. When programming is complete, press Step Function, then Auto Page to return to
the Edit: Auto Run screen.
Parameter Description/Range
Common Vial Carousel number, 1 to 96
Common Vol Volume to withdraw from common vial:
0.0 to 200.0 µL or
0.0 to 2000.0 µL
*
*. Dependent on sample loop installed in Waters 717plus
Autosampler
IDLE
Edit: Auto Addition 1 of 1 Addn
Editing Step:
Run Addition? Yes No
Delay Time (min): 1.0
Common
Vial
1
Step
Function
Delete
Table
Delete
Line
15.0
End of table.
Common
Vol

Table of Contents