Operating the Autosampler 98
4
4.6.3 Using Auto Standards
This section describes:
Overview of Auto Standards
Programming and Running Auto Standards
Example Auto Standard Run
4.6.3.3 Overview of Auto Standards
Auto Standards
Use the Auto Standards routine to program standard vials to run automatically during an
Auto Run.
Use the Auto Standards menu routine to enter standard vial parameters, along with the
number of sample vials to run between standard vials.
The Auto Standards table is saved in permanent memory.
Standard Vials
You can specify up to five standard vials for each Auto Run step. Standard vials:
Can be unique or repeated
Do not have to be in consecutive positions in the carousel
Can be different for each step
Auto Standards Sequence
During an Auto Standards run, the 717plus Autosampler:
Injects standards (specified on Auto Standards screen)
Injects samples (specified on Auto Run screen)
Re-injects the same standards (specified on Auto Standards screen)
Section 4.6.3.5,
Example Auto Standard Run, includes an example of an Auto Standards
run.
4.6.3.4 Programming and Running Auto Standards
Before Programming
Before programming:
Place your samples in the carousel, noting each position (usually in sequential order)