Operating the Autosampler 112
4
.
Figure 4-19 Transfer Before Vial
Transfer Before Injection
If you select Injection as the Transfer sequence, the 717plus Autosampler transfers a
solution to the sample vial before every injection from every vial. Use this method for
multiple (progressive) dilutions of samples.
Figure 4-20
illustrates a one-step Auto Run in which a transfer occurs before each
injection.
Vial
1
Vial
2
Auto Run
Auto Transfer
Step
1
From
2
To
3
Transfer Sequence = Inj
Transfer Vial = 1
Mix = 0
Mix Vol. = 0
Delay Time = 0
Purge Time = 0
Transfer
Inj 1
Inj 2
# Inj
2
Vial
1
Vial
3
Transfer
Inj 1
Inj 2