4
.
Figure 4-20 Transfer Before Injection
4.6.6 Programming the Event Output
The 717plus Autosampler includes a contact closure that you can program to turn on or off
up to four times during each step of an Auto Run. Use the Edit: Step Function screen to
program the contact states and event times.
Note: At the end of an injection, the event is left in the last programmed state. This means
that if an event is on at the end of a run, it will be on at the beginning of the next run,
unless you turn it off.
Vial
1
Vial
2
Auto Run
Auto Transfer
Step
1
From
2
To
3
Transfer Sequence = Inj
Transfer Vial = 1
Mix = 0
Mix Vol. = 0
Delay Time = 0
Purge Time = 0
Transfer
Inj 1
# Inj
2
Vial
1
Vial
3
Transfer
Inj 1
Vial
1
Vial
2
Transfer
Inj 2
Vial
1
Vial
3
Transfer
Inj 2