Startup and Configuration 76
3
3.8.3 Storage
This section covers procedures for long-term storage of the Waters 717plus Autosampler.
If the autosampler will be stored for more than two weeks:
1. Remove buffered mobile phases (see Section 3.8.1,
Removing Buffered Mobile
Phase).
2. Power down the autosampler (see Section 3.8.2,
Powering Down).
3. Remove the waste lines and needle wash line from their containers.
4. Disconnect the red and blue lines.
5. Disconnect the power line.