Running Autosampling 111
4
Figure 4-18 Transfer Before Step
Transfer Before Vial
If you select Vial as the Transfer sequence, the 717plus Autosampler transfers a solution
to the sample vial one time for each vial in the step. The solution is transferred before the
first injection of each vial.
Figure 4-19
illustrates a one-step Auto Run in which the transfer occurs one time for a vial,
then all injections are made from the vial
Vial
1
Vial
2
Vial
3
Vial
4
Auto Run
Auto Transfer
Step
1
2
From
2
7
To
5
10
Transfer Sequence = Step
Transfer Vial = 1
Mix = 0
Mix Vol. = 0
Delay Time = 0
Purge Time = 0
Transfer
Vial
5
Inj Inj Inj Inj
Step 1
Vial
6
Vial
7
Vial
8
Vial
9
Auto Run
Auto Transfer
Step
1
2
From
2
7
To
5
10
Transfer Sequence = Step
Transfer Vial = 6
Mix = 0
Mix Vol. = 0
Delay Time = 0
Purge Time = 0
Transfer
Vial
10
Inj Inj Inj Inj
Step 2