Error Messages 147
6
6.2.4 Heater/Cooler (H/C) Error Messages
6.2.5 Injector (INJ) Error Messages
Table 6-5 H/C Error Messages
Error Message Solution
H/C: Door open time exceeded limit Close door and allow sample compartment
to equilibrate
H/C: High temp limit exceeded Reset autosampler, call Waters
H/C: Internal fan failure Call Waters
H/C: Low temp limit exceeded Reset autosampler, call Waters
H/C: Temp (temperature) control has a
fault
Call Waters
H/C: User aborted defrost cycle Informational message, proceed with
sampling
Table 6-6 INJ Error Messages
Error Message Solution
INJ: Both injector sensors on Reset autosampler, call Waters
INJ: Bottom sensor active Reset autosampler, call Waters
INJ: Compressibility exceeds limit Degas mobile phase. If compressibility test
fails on degassed mobile phase, perform
leak test (Section 6.4.2,
Valve Leak Test).
INJ: Could not find bottom sensor Reset autosampler, call Waters
INJ: Could not find top sensor Reset autosampler, call Waters
INJ: Could not position needle Reset autosampler, call Waters
INJ: Injector seals are not calibrated Calibrate seal pak (Section 3.4.3, Cali-
brating the Seal Pak)
INJ: Pressure inc. (increase) on comp
(compression) check
Perform leak test (Section 6.4.2, Valve
Leak Test)
INJ: Top sensor active Reset autosampler, call Waters