Troubleshooting 148
6
6.2.6 Syringe (SYR) Error Messages
6.3 Troubleshooting Tables
Ta b l e 6 - 8 is a guide to troubleshooting autoinjector problems. It lists autoinjector-related
symptoms, possible causes, and corrective action for the problem.
Table 6-7 SYR Error Messages
Error Message Solution
SYR: Both syringe sensors on Reset autosampler, call Waters
SYR: Cannot move with v2 and v3 closed Open valves 2 and 3 before moving
syringe
SYR: Could not position syringe Reset autosampler, call Waters
SYR: Syringe at empty position Reset autosampler, call Waters
SYR: Syringe at full position Reset autosampler, call Waters
Table 6-8 Autosampler Troubleshooting
Symptom Possible Cause Corrective Action
Autosampler does not
run (fan and front
panel lights off)
Autosampler not
connected to power
source
Ensure power cable is properly
connected to power source and
autosampler.
No power at outlet Check the outlet by connecting another
electrical unit known to be working and
verify if the unit operates. If that unit
does not work, relocate the autosam-
pler to a functioning electrical outlet.