Troubleshooting 144
6
6. Mobile phase(s)
7. Type and model number of detector (UV, RI, fluorescence, conductivity,
electrochemical)
8. Detector settings (wavelength, sensitivity, and so on)
9. Type and serial number of column
10. Sample matrix and components
11. Data system
12. Software version (if applicable)
6.2 Error Messages
This section describes the Waters 717plus Autosampler error messages.
The six error message types provide information about autosampler components and
subsystems (Ta bl e 6 -1
).
Clearing Error Messages
To clear error messages:
Press Clear on the front panel
Correct the source of the error
Table 6-1 Error Message Type
Type Prefix Purpose
Informational Blank or Warning Alerts the operator to incorrect param-
eter entries or operations.
Calibration Error CAL Identifies problems with the injector seal
pak calibration.
Carousel Error CAR Identifies problems with the carousel,
carousel reader, or carousel drive.
Heater/Cooler Error H/C Identifies problems with the
heater/cooler module.
Injector Error INJ Identifies problems with the injector
position sensors.
Syringe Error SYR Identifies problems with the syringe
position sensors.