EasyManua.ls Logo

Waters 717plus - Disconnecting the Sample Loop

Waters 717plus
181 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Changing the Hardware Configuration 53
2
Auxiliary loop from 717plus Autosampler Startup Kit
Union from 717plus Autosampler Startup Kit
To install the loop:
1. From the 717plus front panel, set system flow to 0 mL/min.
2. Turn off the autosampler, disconnect the power cord, and remove the autosampler
cover (see Section 2.6.1,
Removing/Replacing the 717plus Cover).
3. Disconnect the sample loop at the filter using the 5/16-inch wrench (Figure 2-11
).
Figure 2-11 Disconnecting the Sample Loop
4. Connect the auxiliary loop to the filter.
5. Using a union from the Startup Kit, connect the auxiliary loop to the sample loop.
6. Replace the cover, reconnect the power cord, and turn on the autosampler.
7. Complete the configuration procedures discussed in Section 3.6.3,
Entering
Hardware Changes.
Sample Loop
Disconnect at outlet side of
in-line filter.

Table of Contents

Related product manuals