Operating the Autosampler 116
4
The Auto Run Status screen (Figure 4-22) displays:
• Step and Vial Number – Appears in upper right corner, along with a blinking cursor
that indicates that the run is in progress.
• Vial # – Displays current sample vial number.
• Inj – Displays the current injection and number of injections per sample vial on the
current step.
• Volume – Displays the volume injected.
• Runtime – Displays the programmed run time for the current injection.
• Time Remaining – Displays the time remaining for the current injection.
• Total Runtime – Displays the time to run the current and all of the remaining
injections in the Edit: Auto Run table.
• Options – Displays the option that is being run for that step, if any.
Note: Total Runtime does not include the time for executing programmed options,
the sample withdrawal time, or any report delay time.
4.7 Making a Stat Run
The stat run feature allows you to interrupt an auto run to make injections from priority
samples. When the stat run is complete, the autosampler returns to the next injection of
the auto run program.
Performing a Stat Run
To make a stat run:
1. If the autosampler is making an injection, wait until the injection sequence is
complete. Then remove the sample carousel.
2. Load the samples into the carousel in a continuous sequence.
3. From the Auto Run Status screen, press Stat Run, and the Edit Stat Run screen
appears (Figure 4-23
).
Note: The autosampler completes the current injection before beginning the stat run. The
carousel does not release during an injection sequence.