EasyManua.ls Logo

Waters 717plus - 4.5.2 Performing a Purge

Waters 717plus
181 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Ensuring Reproducibility 87
4
4.5.2 Performing a Purge
Perform a purge from the Purge and Compressibility Page.
Purging the Autosampler
To perform a purge using the default values, press the Waters 717plus Autosampler keys
as indicated below (Figure 4-6
):
Figure 4-6 Purging the Autosampler
Changing Purge Time
Change the Purge Time if:
System flow rates are less than 1 mL/min
The auxiliary loop is installed
The purge time setting depends on:
Flow rate of the mobile phase
Sample loop size
To change the Purge Time:
1. From the 717plus Main Page, press Purge Page to display the Purge and
Compressibility page (Figure 4-7
).
TP01951
Main Page
Purge and
CompressibilityPage
Purge
Page
Start
Purge

Table of Contents

Related product manuals