Error Messages 145
6
6.2.1 Informational Error Messages
Informational error messages alert the operator to incorrect parameter entries or
operations (Ta bl e 6 -2
).
6.2.2 Calibration (CAL) Error Messages
Most calibration error messages appear during seal pak calibration (Table 6- 3).
Table 6-2 Informational Error Messages
Error Message Solution
Cannot prg (purge) while initializing Press Clear, wait for initialization to finish
Cannot run stat while purging Press Clear, wait for purge to finish
Inject volume exceeds loop vol (volume) Press Clear, install auxiliary loop for
volumes over 200 µL
Power fail interrupted run Press Clear, and reprogram auto run table
Vial exceeds carousel size Press Clear, check carousel size, and
reenter vial number
Table 6-3 CAL Error Messages
Error Message Solution
CAL: Could not find bottom sensor Recalibrate, call Waters
CAL: Could not find upper seal Recalibrate, call Waters
CAL: Pressure above 750 psi Lower pump pressure (400 to 600 psi),
recalibrate
CAL: Pressure below 400 psi Raise pump pressure (400 to 600 psi),
recalibrate
CAL: Seal width too narrow Recalibrate, call Waters
CAL: Seal width too wide Recalibrate, call Waters
CAL: Top sensor too high Recalibrate, call Waters
CAL: Top sensor too low Recalibrate, call Waters
CAL: Transducer pressure too high Autozero transducer before recalibrating
CAL: Transducer pressure too low Autozero transducer before recalibrating
CAL: User aborted calibration Recalibrate