EasyManua.ls Logo

Waters 717plus - 4.8.2 Microbore;Semiprep Injections

Waters 717plus
181 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operating the Autosampler 120
4
2. Set the sample chamber temperature on Configuration Page 2 by pressing the keys
indicated in Figure 4-25
.
Figure 4-25 Setting Sample Chamber Temperature
3. Close the sample compartment door.
Note: The heater/cooler shuts off if the sample compartment door is left open more
than 5 minutes
4. Wait for the sample compartment to reach the set point. The heater/cooler may
require up to 90 minutes to reach 4° C from an ambient temperature of 27° C.
4.8.2 Microbore/Semiprep Injections
The Waters 717plus Autosampler may be physically reconfigured to facilitate microbore
injections (< 5 µL) and semiprep injections (> 150 µL).
Menu: Main
Configuration
Page 1
Config
Page
NextCfg
Page
IDLE
Edit: Configuration Page 2 of 4
Hold Contact Sense?
Hold Input
Inject Start Pulse (sec) :
End of Rub Pulse (sec) :
Temp Setpoint (
i
C) :
Temp Control?
PrevCfg
Page
+
On
0.5
0.5
8
On
Off
30
No
20
Off
NextCfg
Page
Main
Page
8
On

Table of Contents

Related product manuals